Testo 890 Thermal Imager Firmware 1.66
Dosya tipi: Firmware.
Işleti̇m Si̇stemi: Windows 7, Windows 8, Windows 8.1, Windows Vista, Windows XP.
Dosya adı: fw_t885t890.exe.
Bit: 64 Bit.
Li̇sans: Ücretsiz.
Açıklama
Firmware Güncelleme Prosedürü:
1. *.exe-file download
To update the firmware you need the data of the current version.
2. Unpacking the firmware data
By double-clicking on the downloaded file it will be unzipped.
(Default directory c: \ testoWBK)
3. Preparation of the thermal imager
1) Insert a fully charged battery.
2) Connect the power supply unit with the camera, if the battery isn’t fully charged.
3) Connect imager and computer by using the USB-cable.
4) Switch on imager and wait until the firmware is completely started.
4. Execute the firmware update
1) Run IRSoft and choose Camera/ Configuration.
2) Choose Instrument settings/ Firmware Update …
3) Choose fw_t885t890_Vx.xx.bin and confirm with Open.
The firmware update is now running.
Afterwards the camera will automatically reboot.