Keysight DSOX4032A Oscilloscope Firmware 4.08
Tipo de arquivo: Firmware.
Sistema operacional: AnyOS.
Nome do arquivo: 4000XSeries.04.08.2016071801.agx.
Bit: 32/64 Bit.
Licença: Grátis.
Descrição
Release Notes:
New Features
This software revision includes the following new capabilities:
– Added CXPI triggering and decode option (DSOX4CXPI).
– Added NFC triggering option (DSOX4NFC).
– Added frequency response analysis to Power Application (PSRR and Control Loop Response).
– Added FFT phase math operator.
Enhancements
This software revision includes the following new enhancements:
– Allow all time domain math functions to be loaded into waveform memories.
– Allow waveform memories to be used as sources for math functions.
– Added user-customizable labels for internal setup files.
– Added GUI indicator for channel invert setting.
– Added SCPI grammar for N2820 zoom and R-sense.
– Added option for a custom time reference location.
– Added “no length control” option to LIN 1.3.
– Improved WaveGen duty cycle granularity.
– Extended WaveGen pulse width limits to 1ns at frequencies >=25kHz.
– Extended probe attenuation ratio limits to 0.001 : 1 (-60dB).
– Power Application – Inrush Current – Added support for DC-input power supplies.
– Power Application – Turn-on/Turn-off – Added support for DC-input power supplies.
– Power Application – Turn-on/Turn-off – Added option to specify measurement threshold levels.
– Power Application – Current Harmonics – Added option to automatically determine line frequency.
Bug Fixes
This software revision includes the following Bug fixes:
– Fixed incorrect DVM values in rare circumstances after toggling channel states.
– Fixed error messages when loading CAN FD .dbc files with data fields longer than 8 bytes.
– Fixed crash when saving H5 files with long path strings.
How to Install
– To update your oscilloscope’s firmware, do one of the following:
Place the file on a USB flash drive, connect it to the oscilloscope, press [Utility] > File Explorer, select the file; then, press Load File.
If your oscilloscope is on the network: you can place the file on your computer, access the oscilloscope’s Web Interface (see the User’s Guide for details), click Instrument Utilities, select Firmware Version, browse to select the file; then, click Install.
– After your oscilloscope’s firmware is updated, check the calibration status: press [Utility] > Service > User Cal Status. If “Results: OK” is displayed, you do not need to recalibrate. If calibration is required, press the Start User Cal softkey.
NOTE:
When upgrading your oscilloscope, download the .agx file, and load the complete file into the oscilloscope. Do not “unzip” the file.
Download