ETS-Lindgren FM5004 Field Monitor Firmware 5.4
Dateityp: Firmware.
Betriebssystem: AnyOS.
Dateiname: rev_5_4_firmwareAndNotes.zip.
Bit: 32/64 Bit.
Lizenz: Kostenlos.
Beschreibung
Improvements :
1) Clipped battery percentage values to 100
2) Automatically pick up a probe after a reset from the front panel.
3) Prevent loss of communication with the HI-6005 after a temperature request from the front panel.
4) Cleared OVER RANGE not getting cleared on the averaging screen.
5) Resolved issues with the RDIS command
6) Timeouts on the probe cause timeouts in the FM5004. If these timeouts are not detected at the GPIB level, this can lead to remnants of readings being left in the output GPIB buffer. This causes confusion and it happens several times, can lead to what appears as a lockup.
7) Auto Ranging was improved when the probe was in high power fields
Hinweis:
The PC driver application must be written properly so that the GPIB RDIS,0 command is used to get CXYZ values from the monitor. After receiving a RDIS,0 command, the monitor sends CXYZ values in 4 different lines. The driver must use 4 different reads in order to get each line properly. If a timeout occurs in between any of the reads, the next read will get information that was meant to the last read. For instance, the second read should get an X value. If there is an unhandled timeout, the next read, instead of getting a Y value will get an X value. This problem can lead to out-of-sync RDIS,0 responses where the next RDIS,0 command will not start with a C value, but with the Z value from previous RDIS,0 commands. If enough timeouts occurs, the leftovers can cause the monitor to hang.