Thermo Fisher 46i-HL Analyzer Firmware 02.02.03.304
Tipo de arquivo: Firmware.
Sistema operacional: Windows 7, Windows 8, Windows 8.1, Windows Vista, Windows XP.
Nome do arquivo: 48i020203.zip.
Bit: 64 Bit.
Licença: Grátis.
Descrição
Version 02.02.03 changes relative to version 02.02.00:
1. Fix display of drift compensation for 410.
2. Added concentration calibration data to output list for MODBUS usage.
3. Fix logged data averaging where the sum was divided by ‘n + 1’ instead of ‘n’ seconds, yielding a possible 1.6% error in logged concentration data for 1-minute logging or 0.03% error in hourly concentration data.
4. Fix handling of IP addresses with 3 digits in each of the four fields in INSTRUMENT CONTROLS> COMMUNICATIONS> TCP/IP SETTINGS> IP ADDRESS screen.
5. Fix “set format”, “set lrec format”, “set srec format”, and “set erec format” CLINK commands to save the new setting if the source of the command is serial.
6. Fixed “set program bootloader procboard” CLINK command so that it no longer causes iSeries+ units to hang.
7. The “host name” CLINK command now responds with a <space> if no host name is found. If “set host name” data does not pass validation, it now responds with “data not valid” instead of “bad command”.
UPGRADE FIRMWARE
Below is a procedure for loading the firmware into FLASH memory. The firmware update file transfer process should take about 30 minutes at 57,600 baud. It is assumed that iPort is already talking to the instrument and the instrument window is currently open.
NOTA:
DO NOT TURN OFF THE INSTRUMENT AT ANY TIME DURING THIS UPDATE If the instrument is turned off while burning the new image to the FLASH, it may require replacement of the CPU board, motherboard, I/O expansion board, and/or measurement interface board. To reduce this risk, make sure the instrument is running on clean and stable power before performing this update.
1. Close all instrument windows.
2. From the iPort menu, select Instrument > Update Firmware. Select TCP/IP or Serial, depending on the connection.
3. In the Update Instrument Firmware Program dialog box, enter the instrument ID (if using serial port) or the TCP/IP address (if using TCP/IP).
4. In the Open File dialog box, select the firmware update file, then click the Open button.
5. File transfer progress can be monitored by looking at the transferred blocks in the lower left corner of the iPort window as well as on the instrument’s display.
6. Once the file transfer is complete, the instrument will automatically reboot. There may be some error messages regarding configuration and calibration files that are displayed, this is normal after a firmware update. At this time, the bootloader and application code in each of the low-level processors will be updated to the latest version.
7. To verify all updates were successful, go to the ALARMS menu and make sure the board status alarms at the bottom of the menu all show “OK”. If any board status alarms show “FAIL”, try rebooting the instrument and checking the ALARMS menu again. If they still show “FAIL”, contact technical service.